MIDC/Data Quality Statements
Total Records: 6, displaying 1-6

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSUpwelling LI-20010/15/201000:00:0012/06/201009:01:00The LI-200 is reading zero or bogus during this time. On 10/20/2010, the instrument was removed from service and the channel was shorted. The instrument was disassembled and the Silicon sensor was found to be bad. Another spare sensor was installed on 12/6/2010. Note: This also affects the Albedo (LI-200) calculation.
2BMSUpwelling LI-20010/09/201016:50:0010/14/201010:23:00The LI-200 is reading zero or bogus during this time. On 10/14/2010, the bendix connector at the instrument was completely soldered, which seemed to fix the problem. Note: This also affects the Albedo (LI-200) calculation.
3BMSUpwelling LI-20009/10/201008:50:0010/07/201015:00:00The LI-200 is reading zero during this time. Instrument cable (from logger to bendix connector) was replaced on 10/6/2010, which did not solve the problem. On 10/7/2010, the bendix connector at the instrument was partially re-soldered, which seemed to fix the problem. Note: This also affects the Albedo (LI-200) calculation.
4BMSUpwelling LI-20007/17/201004:00:0009/09/201008:51:00The LI-200 is reading zero during this time. The bendix connect was between the logger and sensor was disconnected, sprayed with contact cleaner, and reconnected, which seemed to fix the problem. Note: This also affects the Albedo (LI-200) calculation.
5BMSUpwelling LI-20007/10/201011:44:0007/14/201013:04:00The LI-200 is reading zero or bogus during this time. The bendix connect was between the logger and sensor was disconnected and reconnected, which seemed to fix the problem. Note: This also affects the Albedo (LI-200) calculation.
6BMSUpwelling LI-20007/03/201000:00:0007/06/201010:03:00The LI-200 is reading zero or bogus during this time due to a loose screw on the multiplexer. Note: This also affects the Albedo (LI-200) calculation.