| R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
| 1 | SRRL BMS | | Photometer TWC | 03/02/2013 | 13:06:00 | 03/02/2013 | 18:24:00 | Due to a scheduled NREL power outage, to upgrade campus power capacity, The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, NIP #1, LI-201, NIP RG780, CUVA2, CUVB2, TWC; and the following instruments were un-shaded: Diffuse 8-48, PSP, CM22, CUV4, PIR, CG4. The Sci-Tec tracker for these insruments are protected by UPS, however the battery depleted at 13:06 MST (the power outage started at 8:06 MST) |
| 2 | SRRL BMS | | Photometer TWC | 03/02/2013 | 08:06:00 | 03/02/2013 | 18:24:00 | Due to a scheduled NREL power outage, to upgrade campus power capacity, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, TSI-880, PGS-100, POM-01, TWC, EFM-100. |
| 3 | SRRL BMS | | Photometer TWC | 07/06/2012 | 19:35:00 | 07/09/2012 | 10:18:00 | A GFCI on the North tracker post was tripped due to thunderstorms. Data from the following instruments are bogus due to not having power during this time: CUVA2, CUVB2, TWC, PGS-100, POM-01. |
| 4 | SRRL BMS | | Photometer TWC | 05/26/2012 | 06:15:00 | 05/26/2012 | 07:12:00 | Due to a planned power outage for SRRL electrical work, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, PGS-100, POM-01, TWC. |
| 5 | SRRL BMS | | Photometer TWC | 11/12/2011 | 11:59:00 | 11/12/2011 | 13:26:00 | Due to an undervoltage followed by a power outage at 13:12, the following BMS instruments may be reading bogus or invalid data: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, TWC. |
| 6 | SRRL BMS | | Photometer TWC | 09/08/2011 | 05:56:00 | 09/08/2011 | 09:38:00 | Due to a planned power outage for SRRL construction activities, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, PGS-100, POM-01, TWC. |
| 7 | SRRL BMS | | Photometer TWC | 07/12/2011 | 22:57:00 | 07/13/2011 | 07:40:00 | Due to a tripped GFCI outlet at SRRL the following instruments were without power during this time: TWC & POM-01 Photometers, Direct CUVA2 & CUVB2, and the PGS100. The readings are either zero or bogus and should not be used. GFCI was tripped due to Eppley venitlators which were water logged by thunderstorm. |
| 8 | SRRL BMS | | Photometer TWC | 06/14/2011 | 13:01:00 | 06/14/2011 | 13:14:00 | Due to a planned power outage for SRRL construction activities, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, TSI-880, PGS-100, POM-01, TWC, EFM-100. |
| 9 | SRRL BMS | | Photometer TWC | 05/25/2011 | 05:19:00 | 05/26/2011 | 12:16:00 | Due to a planned power outage for SRRL construction activities, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, TSI-880, PGS-100, POM-01, TWC, EFM-100. |
| 10 | SRRL BMS | | Photometer TWC | 05/17/2011 | 14:20:00 | 05/17/2011 | 14:48:00 | Due to a planned power outage for SRRL construction activities, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, TSI-880, PGS-100, POM-01, TWC, EFM-100. |
| 11 | SRRL BMS | | Photometer TWC | 05/07/2011 | 05:55:00 | 05/09/2011 | 07:58:00 | There was a planned power outage for SRRL construction activities that started at 5:18 MST. Unfortunately, the backup generator provided to SRRL was providing bad power to the UPS powering the Sci-Tec trackers, therefore the UPS failed at 5:55 MST. Unfortunately, when normal AC power was restored at 11:45 MST, the Sci-Tec trackers had lost all their configuration due to UPS cycling on and off during the outage, therefore were not tracking. The tracker settings were restored on May 9 at 7:58 MST. The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, NIP #1, LI-201, NIP RG780, CUVA2, CUVB2, TWC; and the following instruments were un-shaded: Diffuse 8-48, PSP, CM22, CUV4, PIR, CG4 |
| 12 | SRRL BMS | | Photometer TWC | 05/07/2011 | 05:18:00 | 05/07/2011 | 11:45:00 | Due to a planned power outage for SRRL construction activities, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, TSI-880, PGS-100, POM-01, TWC, EFM-100. |
| 13 | SRRL BMS | | Photometer TWC | 04/15/2011 | 06:11:00 | 04/15/2011 | 06:50:00 | Due to a planned power outage for SRRL construction activities, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210W, SPN1, TSI-880, PGS-100, POM-01, TWC, EFM-100. |
| 14 | SRRL BMS | | Photometer TWC | 03/18/2011 | 00:00:00 | 03/18/2011 | 07:57:00 | The Sci-Tec #2 tracker was jammed on the shading plate. The following instruments were unshaded: CM22, CUV4, CG4; and the following instruments were not pointing at the sun: NIP #1, NIP RG780, LI-201, TWC Photometer. |
| 15 | SRRL BMS | 0001 | Photometer TWC | 10/30/2010 | 00:00:00 | 11/01/2010 | 11:27:00 | The Sci-Tec #2 tracker was apparently not working during this time. The following instruments were unshaded: CM22, CUV4, CG4; and the following instruments were not pointing at the sun: NIP #1, NIP RG780, LI-201, TWC. |
| 16 | SRRL BMS | | Photometer TWC | 12/09/2009 | 12:11:00 | 12/09/2009 | 14:22:00 | Due to a power outage at NREL, followed by bad power, the SRRL tracker UPS shutdown. The following BMS instruments were not pointing at the sun: Direct CH1, NIP1, NIP2, NIP-RG780, LI201, TWC500, CUV4; and the following instruments were unshaded: 8-48, CM22, PSP, CUV4, PIR, CG4. |
| 17 | SRRL BMS | 0001 | Photometer TWC | 07/30/2009 | 12:45:00 | 07/30/2009 | 23:59:00 | The Sci-Tec #2 tracker was apparently not working during this time. The following instruments were unshaded: CM22, CUV4, CG4; and the following instruments were not pointing at the sun: NIP #1, NIP RG780, LI-201. |
| 18 | SRRL BMS | | Photometer TWC | 11/15/2008 | 14:02:00 | 11/15/2008 | 14:32:00 | Due to a planned power outage at SRRL, the following BMS instruments were not pointing at the sun: Direct CH1, NIP1, NIP2, LI201, TWC500, CUV4. |
| 19 | SRRL BMS | 0001 | Photometer TWC | 09/30/2008 | 14:18:00 | 10/01/2008 | 15:00:00 | The TWC photometer was not properly aligned with the sun after being relocated to the Sci-Tec tracker. Shims were added on 10/1/08 to improve alignment. |
| 20 | SRRL BMS | | Photometer TWC | 12/22/2007 | 08:41:00 | 12/22/2007 | 11:24:00 | Due to a scheduled SRRL/Mesa Top power outage on 12/22/2007, the TSP-1, TSP-700, All Global UVA and UVB, and the TWC Photometer were without power, therefore data is either bad or bogus. |
| 21 | SRRL BMS | 0001 | Photometer TWC | 05/25/2006 | 10:50:00 | 05/25/2006 | 11:26:00 | Sun Follower III tracker was down due to power shutdown for installation of new SRRL equipment lift. The following instruments were NOT pointing at the SUN: NIP #1, NIP RG780, CUVB2, LI-201, TWC (500nm). |
| 22 | SRRL BMS | | Photometer TWC | 01/08/2004 | 00:00:00 | 12/21/2005 | 14:53:00 | TWC Photometer was down due to detector failure. Detector and filter was replaced and the unit was returned to service. A new Vo calibration constant will need to be generated on the next clear day. |
| 23 | SRRL BMS | | Photometer TWC | 12/13/2000 | 07:21:00 | 12/15/2000 | 09:46:00 | The 500nm Photometer was disconnected for repair. The instrument had been giving noisy signal at night between midnight and sunrise, and high voltage output (up to 13.5 V) at approx. 7:30 MST (almost every day), which caused the datalogger to put out missing values (-99999) due to over-ranging. This was caused by broken or loose connections in the instrument cable and depending on the position of the tracker the wires supplying power to the instrument shorted with the wires carrying the signal from the instrument. This wire has been repaired are re-oriented to help prevent future damage. |
| 24 | SRRL BMS | | Photometer TWC | 11/01/2000 | | 08/23/2001 | 12:00:00 | An incorrect calibration factor of 0.487 W/mē/nm/mV was in the datalogger during this time period. The correct factor should have been 0.388 W/mē/nm/mV. This data was repaired on 6-14-2002, a correction is NOT required. |