R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
---|---|---|---|---|---|---|---|---|
1 | BMS | 2703 | Global PYR-P | 08/14/2007 | 00:00:00 | 10/20/2008 | 10:37:00 | Instrument is reading 10-20% low for some unknown reason. Signal was verified to be the same on deck and at logger. Instrument may be degrading, however some erratic behavior was also observed on Sept. 3 & 4 when the instrument had several spikes of irradiance levels up to 2500 W/m^2 for periods up to about 15-minutes, then back to "normal". |
2 | BMS | 2703 | Global PYR-P | 03/25/2006 | 00:00:00 | 05/31/2006 | 07:48:00 | Instrument was reading 15-20% high for some unknown reason. Signal was verified to be the same on deck and at logger. Instrument was calibrated in BORCAL 2006-01 and new factor was entered into logger at 7:48 MST on May 31. |