MIDC/Data Quality Statements
Total Records: 4, displaying 1-4

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSGlobal LI-19005/20/202200:00:00The SRRL BMS Global LI-190 Quantum sensor is reading low, most likely due to water contamination into the sensor. It may dry out.
2BMSQ8432Global LI-19009/24/201700:00:0010/16/201700:00:00The SRRL BMS Global LI-190 Quantum sensor is reading about 25% low, problem unknown at this time, but it is most likely water contamination. Signal returned to normal on 10/16/2017, it may have dried out.
3BMSGlobal LI-19004/27/201500:00:00The SRRL BMS Global LI-190 Quantum sensor is reading about 15% low, problem unknown at this time.
4BMSQ8435Global LI-19004/03/200800:00:0004/23/200812:11:00The SRRL BMS Global LI-190 Quantum sensor went faulty on April 3, 2008. The bad sensor (S/N Q8435) was replaced with a spare sensor (S/N Q8432). Sensor started reading 10-20% low starting April 3, was calibrated on April 14, then seemed to shift back to normal, which caused the sensor to read 10-20% high using the new cal factor.