MIDC/Data Quality Statements
Total Records: 4, displaying 1-4

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSGlobal CM6b05/18/202013:03:0012/15/202119:31:00DATA CORRECTED on 12/17/2021: Irradiances were increased by 0.675% due to revised BORCAL report for all BMS shortwave radiometers calibrated in 2020 and 2021.
2BMS015189Global CM6b06/22/201109:00:0006/24/201114:00:00Condensation has formed inside the CM6b dome near the top, which is causing lower irradiance readings around solar noon. The desiccant was replaced which seemed to dry out the dome after a couple days.
3BMS015189Global CM6b06/15/201008:00:0006/29/201016:00:00Condensation has formed inside the CM6b dome near the top, which is causing lower irradiance readings around solar noon. Instrument was opened, dried, and re-sealed between 15:30 to 15:45 MST on 6/29/2010.
4BMS015189Global CM6b05/27/200900:00:0006/05/200910:51:00Condensation has formed inside the CM6b dome near the top, which is causing lower irradiance readings around solar noon. Instrument was removed from service 6/1/2009 10:19 MST, dried, dome and desiccant o-rings replaced, and desiccant replaced. Returned to service 6/5/2009 10:51 MST.