R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | | Global 40-South LI-200 | 05/12/2021 | 10:20:00 | 05/12/2021 | 12:13:00 | The relay multiplexer on the SRRL BMS radiometer tower had a problem (old age/possible moisture) that caused low voltage measurements on the main CR3000 wiring panel to read bogus, this includes all 40-South measurements and all upwelling shortwave measurements (except upwelling CMP11). |
2 | BMS | | Global 40-South LI-200 | 05/14/2020 | 10:38:00 | 12/15/2021 | 19:31:00 | DATA CORRECTED on 12/17/2021: Irradiances were increased by 0.675% due to revised BORCAL report for all BMS shortwave radiometers calibrated in 2020 and 2021. |
3 | BMS | | Global 40-South LI-200 | 12/16/2002 | 14:43:00 | 12/17/2002 | 13:06:00 | The J-Box on the BMS Radiometer Tower was being replaced, data from this instrument is missing or bogus. |