R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | | Downwelling PIR | 09/29/2023 | 12:40:00 | 09/29/2023 | 13:55:00 | Due to an NREL-wide power outage caused by Xcel Energy, The following BMS instruments were not pointing at the sun during this time: Direct CHP-1, sNIP, CUVA2, CUVB2, TUVR; and the following instruments were un-shaded: Diffuse 8-48, CM22-2, PIR. |
2 | BMS | | Downwelling PIR | 08/22/2023 | 13:24:00 | 08/22/2023 | 15:46:00 | Due to BMS south sci-tec failure, The following BMS instruments were not pointing at the sun during this time: Direct CHP-1, sNIP, MS-57, CUVA2, CUVB2, TUVR; and the following instruments were un-shaded: Diffuse 8-48, CM22-2, PIR. However automatic substitution was enabled for the CHP1-1, sNIP, 8-48, and CM22-2. |
3 | BMS | | Downwelling PIR | 08/21/2022 | 05:21:00 | 08/21/2022 | 11:15:00 | Due to BMS south sci-tec failure, The following BMS instruments were not pointing at the sun during this time: Direct CHP-1, sNIP, CUVA2, CUVB2, TUVR; and the following instruments were un-shaded: Diffuse 8-48, CM22-2, PIR. Since the middle sci-tec was down (due to collision with south scitec); data was corrected on 9/14/2022 by subsituting Diffuse 8-48 and CM22-2 with RSR Diffuse, then recalating Direct CHP1-1 and sNIP using BMS Global CMP22 and RSR Diffuse. |
4 | BMS | | Downwelling PIR | 08/20/2022 | 11:56:00 | 08/20/2022 | 23:59:00 | Due to BMS south sci-tec failure, The following BMS instruments were not pointing at the sun during this time: Direct CHP-1, sNIP, CUVA2, CUVB2, TUVR; and the following instruments were un-shaded: Diffuse 8-48, CM22-2, PIR. However automatic substitution was enabled for the CHP1-1, 8-48, and CM22-2. |
5 | BMS | | Downwelling PIR | 06/22/2022 | 16:53:00 | 06/23/2022 | 07:17:00 | Due to BMS south sci-tec failure, The following BMS instruments were not pointing at the sun during this time: Direct CHP-1, sNIP, CUVA2, CUVB2, TUVR; and the following instruments were un-shaded: Diffuse 8-48, CM22-2, PIR. However automatic data substitution was enable for the CHP1-1, 8-48, and CM22-2. |
6 | BMS | | Downwelling PIR | 03/29/2019 | 09:15:00 | 03/31/2019 | 17:00:00 | Heavy rain caused a problem with the PIR signal cable. Data could be up to 30 W/m^2 off. Problem is most likely due to case temperature signal corruption. Cable/connector slowly dried up over the next few days. |
7 | BMS | | Downwelling PIR | 03/15/2019 | 08:00:00 | 03/15/2019 | 16:20:00 | A problem started developing with PIR cable that was causing errors in the PIR case temperature (and the calculated downwelling irradiance), possibly due to melting snow. The PIR irradiance could be up to 30 Watts/m^2 off. Also there will be occasional missing or bogus values form 12:50 to 16:20, when the cable was being troubleshot. A temporary cable has been installed to resolve the issue while the original cable is being repaired. |
8 | BMS | | Downwelling PIR | 12/14/2014 | 10:00:00 | 12/15/2014 | 23:00:00 | There is a problem with the SRRL BMS Downwelling PIR which is causing a possible 10-15 W/m^2 error (too low), it appears to be due to a shift in case temperature reading. Problem could be due to possible water infiltration into temporary J-box which was not oriented in the proper position to shed water. All temporary J-box have been reoriented correctly and PIR connections have seemed to dry out. |
9 | BMS | | Downwelling PIR | 12/28/2013 | 11:55:00 | 12/28/2013 | 15:59:00 | Due to a scheduled NREL power outage, to upgrade campus electrical infrastructure, The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, CUVA2, CUVB2, TUVR; and the following instruments were un-shaded: Diffuse 8-48, PSP, PIR. The South Sci-Tec tracker for these instruments are protected by UPS, however the battery depleted at 11:55 MST (the power outage started at 6:46 MST) |
10 | BMS | | Downwelling PIR | 12/21/2013 | 10:30:00 | 12/21/2013 | 16:22:00 | Due to a scheduled NREL power outage, to upgrade campus electrical infrastructure, The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, CUVA2, CUVB2, TUVR; and the following instruments were un-shaded: Diffuse 8-48, PSP, PIR. The South Sci-Tec tracker for these instruments are protected by UPS, however the battery depleted at 10:30 MST (the power outage started at 7:00 MST) |
11 | BMS | | Downwelling PIR | 03/02/2013 | 13:06:00 | 03/02/2013 | 18:24:00 | Due to a scheduled NREL power outage, to upgrade campus power capacity, The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, NIP #1, LI-201, NIP RG780, CUVA2, CUVB2, TWC; and the following instruments were un-shaded: Diffuse 8-48, PSP, CM22, CUV4, PIR, CG4. The Sci-Tec tracker for these insruments are protected by UPS, however the battery depleted at 13:06 MST (the power outage started at 8:06 MST) |
12 | BMS | | Downwelling PIR | 05/07/2011 | 05:55:00 | 05/09/2011 | 07:58:00 | There was a planned power outage for SRRL construction activities that started at 5:18 MST. Unfortunately, the backup generator provided to SRRL was providing bad power to the UPS powering the Sci-Tec trackers, therefore the UPS failed at 5:55 MST. Unfortunately, when normal AC power was restored at 11:45 MST, the Sci-Tec trackers had lost all their configuration due to UPS cycling on and off during the outage, therefore were not tracking. The tracker settings were restored on May 9 at 7:58 MST. The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, NIP #1, LI-201, NIP RG780, CUVA2, CUVB2, TWC; and the following instruments were un-shaded: Diffuse 8-48, PSP, CM22, CUV4, PIR, CG4 |
13 | BMS | | Downwelling PIR | 05/27/2010 | 18:00:00 | 05/28/2010 | 12:15:00 | Due to improper mounting of the replacement CH1 s/n 080033 the Sci-tec tracker was not tracking properly due to collision with the CH1 against tracker. The following BMS instruments were not pointing at the sun: Direct CH1, NIP2, CUVA2, & CUVB2 and the following instruments were unshaded: 8-48, PSP, PIR. |
14 | BMS | | Downwelling PIR | 12/09/2009 | 12:11:00 | 12/09/2009 | 14:22:00 | Due to a power outage at NREL, followed by bad power, the SRRL tracker UPS shutdown. The following BMS instruments were not pointing at the sun: Direct CH1, NIP1, NIP2, NIP-RG780, LI201, TWC500, CUV4; and the following instruments were unshaded: 8-48, CM22, PSP, CUV4, PIR, CG4. |
15 | BMS | | Downwelling PIR | 11/15/2008 | 14:02:00 | 11/15/2008 | 14:32:00 | Due to a planned power outage at SRRL, the following BMS instruments were unshaded: Diffuse 8-48, CM22, PSP, CUV4, DW-PIR, DW-CG4. |
16 | BMS | 31194F3 | Downwelling PIR | 05/25/2006 | 10:50:00 | 05/25/2006 | 11:10:00 | Sci-Tec tracker #1 tracker was down due to power shutdown for installation of new SRRL equipment lift. The following instruments were UNSHADED: Diffuse 8-48, Diffuse PSP, and the Downwelling PIR. |
17 | BMS | 31194F3 | Downwelling PIR | 01/11/2006 | 14:16:00 | 01/31/2006 | 23:59:00 | The K0 factor was mistakenly entered as -4.8 (instead of the correct -4.58) in the datalogger program. This creates a very minor 0.22 W/m^2 bias in the irradiance levels. |
18 | BMS | 31192F3 | Downwelling PIR | 07/23/2005 | 10:16:00 | 07/23/2005 | 17:22:00 | Sci-Tec tracker #1 was down due to a scheduled NREL power outage on 7/23/2005 at 5:33 MST. The following instruments were UNSHADED: Diffuse 8-48, Diffuse PSP, and the Downwelling PIR. |
19 | BMS | 31192F3 | Downwelling PIR | 07/10/2005 | 09:30:00 | 07/10/2005 | 23:59:00 | The SRRL BMS Sci-Tec Tracker #1 was hung up on cables during this time. The following instruments were UNSHADED: Diffuse 8-48, Diffuse PSP and the PIR. |
20 | BMS | | Downwelling PIR | 10/30/2003 | 04:00:00 | 11/07/2003 | 21:00:00 | A problem with the instrument cable was causing the PIR Case Temperature to read approx. 2°C high, which caused the downwelling infrared to read about 50 W/m² high. The downwelling IR is also used for calculating the following: Net Radiation (Eppley), Net Infrared PIR (atmos), and the Diffuse PSP (vent/thermal corrected). |
21 | BMS | | Downwelling PIR | 01/14/1999 | 11:15:00 | 01/20/1999 | 10:50:00 | Downwelling and Upwelling Infrared (PIRs) were not operating correctly during this time frame, this data cannot be fixed. |