MIDC/Data Quality Statements
Total Records: 17, displaying 1-17

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1SRRL BMSDirect TUVR03/15/200400:00:0012/31/200523:59:00The Direct TUVR is experiencing intermittent signal problems, some days its good others its bad. Instrument was permanently removed from service sometime after 6/30/06, however data after 12/31/05 is not available (or should be ignored).
2SRRL BMSDirect TUVR12/27/200300:00:0003/10/200410:11:00The Direct TUVR is experiencing intermitten signal problems, some days its good others its bad. On 3-10-2004, the TUVR sensor was shorted from 9:42 to 10:10 MST in an effort to reset the detector.
3SRRL BMSDirect TUVR08/12/200305:00:0009/26/200318:00:00The Direct TUVR is experiencing intermitten signal problems, some days its good others its bad.
4SRRL BMSDirect TUVR03/13/200200:00:0006/07/200208:08:00The Direct UV is experiencing intermitten signal problems, some days its good others its bad. The instrument was factory serviced and the problem has been corrected.
5SRRL BMSDirect TUVR12/02/200110:00:0002/19/200215:01:00The Direct UV is experiencing intermitten signal problems, some days its good others its bad. On 2-19-2002, the TUVR sensor was shorted from 13:00 to 15:00 in an effort to reset the detector. This seemed to solve the problem.
6SRRL BMSDirect TUVR06/27/200109:46:0008/08/200110:10:00The direct normal UV was removed from service due to moisture leak. A new fused silica glass window and mount has been fitted by the machine shop to avoid future moisture leaks.
7SRRL BMSDirect TUVR05/29/200114:01:0006/22/200115:30:00The direct normal UV was removed from service due to moisture leak.
8SRRL BMSDirect TUVR12/13/200013:18:0012/14/200015:00:00The direct normal UV was down due to broken signal cable.
9SRRL BMSDirect TUVR11/01/200000:00:0011/06/200013:03:00The direct normal UV is disconnected, instrument is awaiting special mount for Sci-Tec tracker.
10SRRL BMSDirect TUVR07/07/200011:00:0007/10/200007:45:00The direct normal UV radiometer was down due to tracker failure.
11SRRL BMSDirect TUVR06/21/200009:30:0006/23/200013:24:00The direct normal UV radiometer was down due to tracker failure.
12SRRL BMSDirect TUVR06/01/200008:30:0006/01/200010:50:00The direct normal UV radiometer was down due to tracker failure.
13SRRL BMSDirect TUVR04/23/200004/24/200008:17:00The direct normal UV radiometer was down due to tracker failure.
14SRRL BMSDirect TUVR04/11/200004/12/200008:46:00The direct normal UV radiometer was down due to tracker failure caused by a power failure that occurred on April 11.
15SRRL BMSDirect TUVR10/06/199914:00:0010/10/199913:40:00The direct normal UV radiometer was removed from the solar tracker and the signal shorted during the cavity comparisons.
16SRRL BMSDirect TUVR03/03/199903/31/199914:44:00The Direct Normal Total Ultraviolet Radiometer (TUVR) s/n 18115 had an intermittent signal problem and was removed for service during this time period. The instrument was sent to The Eppley Laboratory, Inc. for repair and calibration.
17SRRL BMSDirect TUVR09/04/199813:30:0009/17/199809:20:00Direct Normal Ultraviolet radiometer was disabled after solar tracker failure. We installed a signal shorting plug (output fixed at 0.00 VDC).