MIDC/Data Quality Statements
Total Records: 2, displaying 1-2

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMS25782F3Global PSP02/28/200607:58:0002/28/200609:07:00The SRRL BMS Global PSP was also removed from service, so that the instrument's RSnet could be characterized in the NREL Black Body System. It's output is set to missing (-99999).
2BMSGlobal PSP08/01/200000:00:0010/31/200023:59:00SRRL BMS was undergoing upgrade and reinstallation on the new instrument deck during this time. To keep a continuos data set for BMS, Global PSP data from SRRL BSRN was used. Note: BSRN global data was 7.872% low (based on BMS/BSRN comparison in 7-2000), the data was corrected before adding to BMS.