R# | Serial Number | Calibration Facility | Calibration Date | Date for next Calibration | Responsivity [45] | Calibration Factor [45] | Instrument::Make | Instrument::Model | Instrument::Description | Instrument::Program | Instrument::Factory Calibration | Instrument::Date of Factory Calibration | Responsivity Bins [2 deg] | Responsivity Function Coefficients | Other Calibrations | Calibration File | Contact Person | Documentation Title | Calibration Method | Calibration Notes | Other Information | Datalogger Entry Date | Responsivity [45] [Corr] | Calibration Factor [45] [Corr] | Correction Method | Responsivity Bins Average [2 deg] |
1 | 29238 | NREL/Optics Lab | 05/16/1996 | 05/16/1997 | 4.46 W/sm/mV | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | Daryl Myers | see paper file | | uncert=+/-15% | | | | | | |
2 | 29238 | NREL/Optics Lab | 08/28/1996 | 08/28/1997 | Cf=4.461 W/sm/mv | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | D Myers | | | | | | | | | |
3 | 29238 | NREL | 01/06/2000 | 01/06/2001 | 3.350 W/m^2/mV | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | Afshin Andreas | Calibration Report | Outdoors using OL754 | | | | | | | |
4 | 29238 | NREL/Optics Lab | 12/12/2013 | 04/12/2014 | CF = 5.5 W/m²/mV | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | Afshin Andreas | | 280-400nm vs BMS Calculated Direct Total UV (Global & Diffuse CUV4) at noon | TUVR w/ Custom Direct Tube & Window | SRRL BMS UV Radiometer | 12/12/2013 | | | | |
5 | 29238 | NREL | 05/28/2014 | 05/28/2015 | CF = 5.247 W/m²/mV | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | [BMS_DirUVT_2012_06.pdf] | | Afshin Andreas | SRRL BMS UV-Total Calibration Report | 280-400nm vs OL756 w/direct tube | TUVR w/ NREL custom direct tube & window | SRRL BMS UV Radiometer | 06/03/2014 | | | | |
6 | 29238 | NREL | 07/14/2016 | 07/14/2016 | CF = 5.092 W/m²/mV | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | Afshin Andreas | SRRL BMS UV-Total Calibration Report | 280-400nm vs OL756 w/direct tube | TUVR w/ NREL custom direct tube & window | SRRL BMS UV Radiometer | 07/14/2016 | | | | |
7 | 29238 | NREL | 07/31/2018 | 07/31/2019 | CF = 5.099 W/m²/mV | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | Afshin Andreas | | 280-400nm vs OL756 w/direct tube | TUVR w/ NREL custom direct tube & window | SRRL BMS UV Radiometer | 08/02/2018 | | | | |
8 | 29238 | NREL | 07/31/2020 | 07/31/2021 | CF = 4.962 W/m²/mV | | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | Afshin Andreas | | 280-400nm vs OL756 w/direct tube | TUVR w/ NREL custom direct tube & window | SRRL BMS UV Radiometer | 08/03/2020 | | | | |
9 | 29238 | NREL | 07/08/2022 | 07/08/2023 | | CF = 4.898 W/m²/mV (280-400nm) | Eppley | TUVR | Ultraviolet Radiometer | OPTIC | | | | | | | Afshin Andreas | | vs OL756 w/Direct Tube Outdoors | Uncert = ±6% (outdoors) | SRRL BMS UV Radiometer | 07/08/2022 | | | | |