MIDC/AIM Calibration Database
Total Records: 4, displaying 1-4

R#Serial NumberCalibration FacilityCalibration DateDate for next CalibrationResponsivity [45]Calibration Factor [45]Instrument::MakeInstrument::ModelInstrument::DescriptionInstrument::ProgramInstrument::Factory CalibrationInstrument::Date of Factory CalibrationResponsivity Bins [2 deg]Responsivity Function CoefficientsOther CalibrationsCalibration FileContact PersonDocumentation TitleCalibration MethodCalibration NotesOther InformationDatalogger Entry DateResponsivity [45] [Corr]Calibration Factor [45] [Corr]Correction MethodResponsivity Bins Average [2 deg]
1010284-UW-CG3Kipp & Zonen02/16/200102/16/2002RS=9.500 V/W/mCF=105.26 W/m/mVKipp & ZonenCG3-UWNet RadiometerNREL/ProgramCNR1 Factory CalibrationAll (4) instruments were adjusted to same RS08/04/2001
2010284-UW-CG3NREL01/19/200601/19/2007K1=0.093 W/m/V; K2=1.005Kipp & ZonenCG3-UWNet RadiometerNREL/ProgramAfshin Andreasvs BMS UW PIR 31192F3 (1/13/06 through 1/19/06)U95 wrt WISG = 3.8 W/musing Mfg. equation for calculating infrared irradiance01/20/2006
3010284-UW-CG3NREL01/19/200601/19/2007K0=0; K1=0.0895 W/m/V; K2=1.005; K3=0; Kr=7.044e-4Kipp & ZonenCG3-UWNet RadiometerNREL/ProgramAfshin Andreasvs BMS UW PIR 31192F3 (1/13/06 through 1/19/06)U95 wrt WISG = 3.0 W/m02/01/2006
4010284-UW-CG3NREL04/17/200804/17/2009K0=0; K1=0.093 W/m/V; K2=1.0095; K3=0; Kr=7.044e-4Kipp & ZonenCG3-UWNet RadiometerNREL/ProgramAfshin Andreasvs BMS UW PIR 31194F3 (4/9/08 through 4/17/08)U95 wrt WISG = 4.95 W/m04/28/2008