R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
26 | BMS | 0212-2 | Global TSP-700 | 07/14/2009 | 00:00:00 | 11/11/2009 | 12:30:00 | There is an occasional drastic noise or signal drop out problem with the SRRL BMS TSP-700 during this time. Some days there is no problem. Several screws in the power supply box were retightened which seemed to fix the problem. |
27 | BMS | 0212-2 | Global TSP-700 | 11/15/2008 | 07:45:00 | 11/15/2008 | 14:29:00 | Due to a planned power outage at SRRL, the following BMS instruments were down: TSP-1, TSP-700, UVSAT, UVSBT, CUVA1, CUVA2, CUVB1, CUVB2, UVB-1, 501A, MS210, SPN1, Snow Depth, TSI-880, SkyCam. |
28 | BMS | 0212-2 | Global TSP-700 | 09/26/2008 | 00:00:00 | 03/04/2009 | 14:00:00 | There is an occasional drastic noise problem with the SRRL BMS TSP-700, which started showing up 9/26/2008. Every screw between the data logger and instrument was retightened, this includes the patch panel and power supply box, which seemed to solve the problem. |
29 | BMS | 0212-2 | Global TSP-700 | 05/29/2008 | 00:00:00 | 06/09/2008 | 13:00:00 | The signal from the TSP-700 would occasionally drop low during this time frame. A screw was tightened in the power supply box, which seemed to fix the problem. |
30 | BMS | 0212-2 | Global TSP-700 | 12/22/2007 | 08:41:00 | 12/22/2007 | 11:24:00 | Due to a scheduled SRRL/Mesa Top power outage on 12/22/2007, the TSP-1, TSP-700, All Global UVA and UVB, and the TWC Photometer were without power, therefore data is either bad or bogus. |
31 | BMS | 0212-2 | Global TSP-700 | 05/25/2006 | 10:50:00 | 05/25/2006 | 11:10:00 | Data from the Yankee instrument is bad or bogus due to power shutdown for installation of new SRRL equipment lift. |
32 | BMS | 0212-2 | Global TSP-700 | 11/18/2005 | 08:20:00 | 11/18/2005 | 12:06:00 | The SRRL BMS Yankee TSP-700 was down for power supply upgrade. The old Condor HAA15-0.8-A+ (+/-12 V @ 1 Amp) was replaced with a new Condor HB12-1.7-A+ (+12 V @ 1.7 Amp). The new higher current power supply can now restart the TSP-700 after a power failure. Note: The -12 V output on the old power supply was not needed (Yankee had designed it as a universal power supply for all their instruments). |
33 | BMS | 0212-2 | Global TSP-700 | 11/04/2005 | 23:14:00 | 11/09/2005 | 08:42:00 | The SRRL BMS Yankee TSP-700 has been down due to a design flaw or fault with the Yankee power supply, caused by a power outage on 11/4/2005 at 23:14 MST. The power supply is unable to power up both the ventilator and the heater simultaneously (even though it was designed to). Data during this time is bogus. |
34 | BMS | 0212-2 | Global TSP-700 | 07/23/2005 | 05:33:00 | 07/25/2005 | 07:18:00 | The SRRL BMS Yankee TSP-700 has been down due to a design flaw or fault with the Yankee power supply, caused by a scheduled NREL power outage on 7/23/2005 at 5:33 MST. The power supply is unable to power up both the ventilator and the heater simultaneously (even though it was designed to). Data during this time is bogus. |
35 | BMS | 0212-2 | Global TSP-700 | 02/16/2005 | 09:02:00 | 02/17/2005 | 08:51:00 | The SRRL BMS Yankee TSP-700 has been down due to a design flaw or fault with the Yankee power supply, caused by a scheduled SRRL power outage on 2/16/2005 at 9:02 MST. The power supply is unable to power up both the ventilator and the heater simultaneously (even though it was designed to). Data during this time is bogus. |